Next: Results
Up: EOSC252 Lab2: Porosity Measurment
Previous: Introduction
Be sure to record exactly what you do, what readings you take, what
valves you open or close at various stages in this experiment. THINK
ABOUT WHAT PARTS OF THE SYSTEM ARE CONNECTED AT ALL TIMES.
****DANGER!!!! NEVER OPEN THE SYSTEM WITHOUT FIRST LETTING OUT
THE GAS SO THE PRESSURE DROPS TO ZERO.****
- Measure the length and the diameter of your sample. Repeat the 5
times.
|
1 |
2 |
3 |
4 |
5 |
Average |
Diameter () |
|
|
|
|
|
|
|
|
|
|
|
|
|
Length () |
|
|
|
|
|
|
|
|
|
|
|
|
|
- Determine the volume of the sample cell,
, and
expansion cell,
, of the porosimeter
- Ensure V1 is closed tight. Open V2 and V3. Wait until the
reading on the multimeter is stable. Record the voltage
corresponding to the atmospheric pressure.
- Ensure V2 and V3 are closed tight. Open V1 to let gas
flow into sample cell only. When reading on multimeter stabilize,
close V1. Record reading.
- Ensure V1 and V2 are closed tight. Open V3 to let the gas
flow to the expansion cell. Record reading.
- Open V2 to release the gas from the system. Make sure V3 is
open. Close V3.
- Measure the diameter of the calibration sphere using the
caliper. Take 5 measurements.
|
1 |
2 |
3 |
4 |
5 |
Average |
Diameter () |
|
|
|
|
|
|
|
|
|
|
|
|
|
- Ensure that V1 is shut tight and that no gas is left in the
system. Open the sample cell by unscrewing the lid and place the
calibration sphere inside. Close the lid properly.
- Repeat the steps (b) through (d).
- Calculate the volume of the sample cell,
, and of the
expansion cell,
. Make sure you have a clean account of
your calculations. Don't forget to correct for the atmospheric
pressure. If you set the problem right, you should end up with
2 equations and 2 unknowns that are
and
. Make
sure that your solution is an expression of a volume by looking at
the unit and verify that the order of magnitude is consistent with
the size of the components of the porosimeter (e.g. a volume of
would be excessive).
- Determine the porosity of your rock sample
- Ensure that V1 is shut tight and that no gas is left in the
system. Open the sample cell by unscrewing the lid and place your
sample inside. Close the lid properly.
- Ensure V2 and V3 are closed tight. Open V1 to let gas
flow into sample cell only. When reading on multimeter stabilize,
close V1. Record reading.
- Ensure V1 and V2 are closed tight. Open V3 to let the gas
flow to the expansion cell. Record reading.
- Open V2 to release the gas from the system. Make sure V3 is
open. Close V3.
- Calculate the effective volume of your sample (i.e. the volume
that is not filled with helium) using
and
. Now
use this value to find the its porosity. Before you do your
calculations, think about which volume you are measuring. Are all
the pores filled with helium? What is the difference between water
and helium? Is the maximum saturation different when using helium
instead of water? Please comment on that in your discussion.
Next: Results
Up: EOSC252 Lab2: Porosity Measurment
Previous: Introduction
Felix Herrmann
2007-01-19